Journal article
Phantoms for Cross-Calibration of Dual Energy X-ray Absorptiometry Measurements in Infants
Abstract
OBJECTIVE: To test the suitability of phantoms to cross-calibrate body composition measurements in small subjects among different dual energy X-ray absorptiometry (DXA) instruments.
METHODS: A set of four phantoms with total weights 1520g, 3140g, 4650g and 7490g were made with low cost and easily available materials. Each phantom was made from assembling polyethylene bottles (100 to 1000 mL) filled with either pure olive oil or electrolyte …
Authors
Hammami M; Picaud J-C; Fusch C; Hockman EM; Rigo J; Koo WWK
Journal
Journal of the American Nutrition Association, Vol. 21, No. 4, pp. 328–332
Publisher
Taylor & Francis
Publication Date
8 2002
DOI
10.1080/07315724.2002.10719230
ISSN
2769-7061