Some infant ventilators do not limit peak inspiratory pressure reliably during active expiration

Abstract

In order to minimize barotrauma in newborn infants with respiratory failure, peak inspiratory pressures should not exceed those required for adequate gas exchange. We examined whether four commonly used pressure-limited, constant flow ventilators limit pressure reliably during simulated active expiration against the inspiratory stroke of the ventilator. Three machines of each type were tested at 13 different expiratory flow rates (2 to 14 …

Authors

KIRPALANI H; SANTOS-LYN R; ROBERTS R

Journal

Critical Care Medicine, Vol. 16, No. 9, pp. 880–883

Publisher

Wolters Kluwer

Publication Date

9 1988