Evaluation of Cobas Integra® 800 under Simulated Routine Conditions in Six Laboratories

Abstract

The new selective access analyser Cobas Integra 800 from Roche Diagnostics was evaluated in an international multicentre study at six sites. Routine simulation experiments showed good performance and full functionality of the instrument and provocation of anomalous situations generated no problems. The new features on Cobas Integra 800, namely clot detection and dispensing control, worked according to specifications. The imprecision of Cobas …

Authors

Redondo FL; Bermudez P; Cocco C; Colella F; Graziani MS; Fiehn W; Hierla T; Lemoël G; Belliard A; Manene D

Journal

Clinical Chemistry and Laboratory Medicine, Vol. 41, No. 3, pp. 365–381

Publisher

De Gruyter

Publication Date

March 2003