Home
Scholarly Works
Morphine overdose from error propagation on an...
Journal article

Morphine overdose from error propagation on an acute pain service

Abstract

PurposeTo highlight a case in which multiple errors occurred during programming and administration of analgesia via a patient-controlled analgesia (PCA) pump, and to formulate recommendations on how to avoid such errors in the future.Clinical featuresFollowing lumbar surgery, a 43-yr-old woman was switched from epidural analgesia to a PCA pump. This change was associated with numerous errors at several points of delivery of her care. Errors included incorrect connection of the PCA adapter, incorrect pump programming, and communication lapses which resulted in a morphine overdose and subsequent respiratory arrest. The patient was promptly resuscitated, and she had an uneventful recovery. The event resulted in a complete review of pain management equipment and the training and education of staff using this equipment at our institution.ConclusionThis case highlights how multiple individual errors can combine to result in a serious adverse event. While equipment design was an important factor in this adverse event, human factors played a critical role at multiple levels.

Authors

Syed S; Paul JE; Hueftlein M; Kampf M; McLean RF

Journal

Journal canadien d'anesthésie, Vol. 53, No. 6,

Publisher

Springer Nature

Publication Date

January 1, 2006

DOI

10.1007/bf03021849

ISSN

0832-610X