Home
Scholarly Works
Computerized Cheirometer for Assessing...
Journal article

Computerized Cheirometer for Assessing Grip-Strength Indices and Classification of Rheumatoid Patients

Abstract

A computerized Cheirometer has been developed to objectively assess grip strength indices (GSI) of normals and rheumatoid arthritis (RA) patients. The instrument consists of a hand-grip bladder, an LVDT pressure transducer (Schaevitz) interfaced through a 12 bit A/D converter to an Apple II + computer. Software has been developed for data acquisition, display and classification of measured GSI into moderately diseased and severely diseased and normal classes. During the developmental phase of the instrument, it was noted that maximum pressure ( P M ) and rate-of-rise (ROR) of pressure during the squeezing of the bladder were the most sensitive indices to distinguish and classify normal and patient hand-grip waveforms. The reproducibility of P M and ROR as useful GS indices was tested in 13 chronic RA patients in the morning, with two measurements per patient, within half-hour intervals. The results indicate that P M and ROR are stable indices with no significant variation between two measurements ( p < 0.05). A Bayes classifier program was developed to recognize and classify normal and pathological signatures of grip strength. The classifier accuracy was evaluated in 79 patients in the Rheumatology clinic at McMaster Hospital. A classification accuracy of 90 per cent was obtained in classifying normals and pathological subjects.

Authors

Chan APK; Kamath MV; Buchanan WW; Ghista DN; Grace EM; Anvari P

Journal

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine, Vol. 17, No. 1, pp. 17–22

Publisher

SAGE Publications

Publication Date

January 1, 1988

DOI

10.1243/emed_jour_1988_017_007_02

ISSN

0954-4119
View published work (Non-McMaster Users)

Contact the Experts team