Home monitoring report from a single lead Lumax DX implantable cardioverter defibrillator: New observations in a new system

Abstract

A 56-year-old man underwent a single lead Lumax 640 DX implantable cardioverter defibrillator implantation for primary prevention of sudden cardiac death. A DX system consists of a single lead, which provides atrial as well as ventricular electrograms, and enhances atrial arrhythmia detection. Three months after the implantation, high-frequency episodes were detected on the far field and the atrial channels, but not on the bipolar right …