Home
Scholarly Works
Quantifying Stranded Implant Displacement...
Chapter

Quantifying Stranded Implant Displacement Following Prostate Brachytherapy

Abstract

We aim to compute radioactive stranded-implant displacement during and after prostate brachytherapy. We present the methods used to identify corresponding seeds in planned, intra-operative and post-implant patient data that enable us to compute seed displacements. A minimum cost network flow algorithm is used, on 8 patients, for needle track detection to group seeds into needles that can be matched between datasets. An iterative best line detection algorithm is used both to help with needle detection and to register the different datasets. Our results show that there was an average seed misplacement of 5.08±2.35 mm during the procedure, which then moved another 3.10±1.91 mm by the time the quality assurance CT was taken. Several directional trends in different regions of the prostate were noted and commented on.

Authors

Lobo J; Moradi M; Chng N; Dehghan E; Fichtinger G; Morris WJ; Salcudean SE

Book title

Medical Image Computing and Computer-Assisted Intervention – MICCAI 2011

Series

Lecture Notes in Computer Science

Volume

14

Pagination

pp. 307-314

Publisher

Springer Nature

Publication Date

October 11, 2011

DOI

10.1007/978-3-642-23623-5_39
View published work (Non-McMaster Users)

Contact the Experts team