A high resolution, high counting rate, bidimensional MWPC imaging detector for small angle X-ray diffraction studies

Abstract

The performance is reported of a 200 mm × 200 mm X-ray imaging MWPC aimedat applications in small angle X-ray diffraction and scattering. With quantum energies of ∼ 8 keV we report high spatial resolution (±0.5 mm x ±0.14 mm) with a capability for data taking at ⪆350 kHz. The detection efficiency is ∼75% and the detector operates as a sealed unit with a long lifetime.

Authors

Bateman JE; Connolly JF; Sawyer EC; Stephenson R

Journal

Nuclear Instruments and Methods in Physics Research, Vol. 196, No. 2-3, pp. 515–520

Publisher

Elsevier

Publication Date

May 1, 1982

DOI

10.1016/0029-554x(82)90125-2

ISSN

0167-5087