Evaluation of geometries appropriate for 125I in vivo bone strontium X-ray fluorescence measurement
Abstract
In this study, we present a significant improvement in the minimum detection limit (MDL) for strontium measurement in bone with X-ray fluorescence using 125I as the excitation source. The improvement was achieved thanks to a new collimator in the 90 degree geometry and a different source positioning in the 180 degree geometry. The achieved MDL in the 90 degree geometry is 0.0426±0.0006 mg Sr/g Ca, which is 5.5 times better than the one …