Journal article
A primary method for determination of optical parameters of turbid samples and application to intralipid between 550 and 1630 nm.
Abstract
The lack of a primary method for determination of optical parameters remains a significant barrier in optical study of turbid media. We present a complete system of experimental setups and Monte Carlo modeling tools for fast and accurate solution of the inverse problem from the measured signals of homogeneous turbid samples. The calibration of the instrument and validation of the Monte Carlo modeling have been carried out to ensure the accuracy …
Authors
Chen C; Lu JQ; Ding H; Jacobs KM; Du Y; Hu X-H
Journal
Optics Express, Vol. 14, No. 16, pp. 7420–7435
Publisher
Optica Publishing Group
Publication Date
2006
DOI
10.1364/oe.14.007420
ISSN
1094-4087