Journal article
Neutralization—reionization experiments with a cell containing potassium vapour
Abstract
The design, operation and performance of a potassium vapour cell for exothermic neutralization—reionization (NR) experiments using a VG-ZAB-2F mass spectrometer is described. With the new cell it is possible to maintain over a long period of time a sufficiently high and very stable vapour pressure and this enables the acquisition of many noise-free exothermic NR spectra as single scans. Also with the new cell there is no or minimal leakage of …
Authors
Burgers PC; Kulik W; Versluis C; Terlouw JK
Journal
International Journal of Mass Spectrometry, Vol. 98, No. 3, pp. 247–258
Publisher
Elsevier
Publication Date
August 1990
DOI
10.1016/0168-1176(90)80005-n
ISSN
1387-3806