Label-free biosensor array based on silicon-on-insulator ring resonators addressed using a WDM approach.

Abstract

We report a silicon-on-insulator ring resonator biosensor array with one output port, using wavelength division multiplexing as the addressing scheme. With the use of on-chip referencing for environmental drift cancellation, simultaneous monitoring of multiplexed molecular bindings is demonstrated, with a resolution of 0.3 pg/mm(2) (40 ag of total mass) for protein concentrations over 4 orders of magnitude down to 20 pM. Reactions are measured over time periods as long as 3 h with high stability.

Authors

Xu D-X; Vachon M; Densmore A; Ma R; Delâge A; Janz S; Lapointe J; Li Y; Lopinski G; Zhang D