Safety of inhaled corticosteroids delivered by plastic and metal spacers Academic Article uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • BACKGROUND: Because of its non-electrostatic properties the metal Nebuchamber (NC) valved holding chamber (VHC) delivers a greater mass of aerosol to the mouth than the polypropylene Aerochamber (AC) VHC. Delivery of more aerosol to the lungs may also increase systemic absorption of inhaled corticosteroids (ICS) and hypothalamo-pituitary-adrenal (HPA) suppression. METHODS: Thirty children (mean 4.3 (SD 0.3) years) received 200 micro g budesonide twice daily by NC or AC, both with the mask provided, in a randomised, two month crossover trial. Twenty four hour urinary free cortisol (UFC) was determined as a measure of HPA suppression. RESULTS: UFC decreased from 42.3 (7.8) nmol UFC/nmol creatinine control to 26.2 (2.4) (p = 0.06 v control) after AC, and to 24.5 (2.5) (p = 0.04 v control) after NC (p = 0.4 AC v NC). CONCLUSIONS: Despite a greater total dose delivered to the mouth, NC is not associated with greater HPA suppression when using 400 micro g/day budesonide under real life conditions in young children.

publication date

  • June 1, 2003

has subject area