A Comparison of Lyophilized Artificially Depleted Plasmas and Lyophilized Plasmas From Patients Receiving Warfarin in Correcting for Coagulometer Effects on International Normalized Ratios

Abstract

The comparative value of lyophilized artificially depleted plasmas and plasmas from warfarin-treated patients in local (thromboplastin/coagulometer) system International Sensitivity Indices (ISI) determination has been studied at 39 centers in an international collaborative study. At each center, both types of plasmas were tested with two common thromboplastins (low ISI RecombiPlasTin and high ISI OBT), and in-house reagents using three brands …