TIGRESS highly-segmented high-purity germanium clover detector

Abstract

The TRIUMF-ISAC Gamma-Ray Escape-Suppressed Spectrometer (TIGRESS) will consist of twelve units of four high-purity germanium (HPGe) crystals in a common cryostat. The outer contacts of each crystal will be divided into four quadrants and two lateral segments for a total of eight outer contacts. The performance of a prototype HPGe four-crystal unit has been investigated. Integrated noise spectra for all contacts were measured. Energy …

Authors

Scraggs HC; Pearson CJ; Hackman G; Smith MB; Austin RAE; Ball GC; Boston AJ; Bricault P; Chakrawarthy RS; Churchman R

Journal

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment, Vol. 543, No. 2-3, pp. 431–440

Publisher

Elsevier

Publication Date

5 2005

DOI

10.1016/j.nima.2004.12.012