Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Monte Carlo design study for thick gas electron...
Journal article

Monte Carlo design study for thick gas electron multiplier-based multi-element microdosimetric detector

Abstract

To accomplish enhanced neutron dose response with high detection efficiency, a set of multi-element microdosimetric detectors were designed using THick Gas Electron Multiplier (THGEM). THGEM generates a strong electric field within microholes of a sub-millimeter thick insulator, which makes electron multiplication possible without the traditional anode wire electrodes. Owing to the absence of wire electrodes, the newly designed neutron …

Authors

Anjomani Z; Hanu AR; Prestwich WV; Byun SH

Journal

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment, Vol. 757, , pp. 67–74

Publisher

Elsevier

Publication Date

9 2014

DOI

10.1016/j.nima.2014.04.063

ISSN

0168-9002