A Colloidal Stability Assay Suitable for High-Throughput Screening Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • A library of 32 polystyrene copolymer latexes, with diameters ranging between 53 and 387 nm, was used to develop and demonstrate a high-throughput assay using a 96-well microplate platform to measure critical coagulation concentrations, a measure of colloidal stability. The most robust assay involved an automated centrifugation-decantation step to remove latex aggregates before absorbance measurements, eliminating aggregate interference with optical measurements made through the base of the multiwell plates. For smaller nanoparticles (diameter <150 nm), the centrifugation-decantation step was not required as the interference was less than with larger particles. Parallel measurements with a ChemiDoc MP plate scanner gave indications of aggregation; however, the results were less sensitive than the absorbance measurements.

authors

  • Abarca, Carla
  • Ali, M Monsur
  • Yang, Songtao
  • Dong, Xiaofei
  • Pelton, Robert

publication date

  • March 1, 2016