In vitro assessment of the biocompatibility of chemically treated silicone materials with human lens epithelial cells Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • AbstractCytotoxicity testing is a regulatory requirement for safety testing of new ocular implants. In vitro toxicity tests determine whether toxic chemicals are present on a material surface or leach out of the material matrix. A method of evaluating the cytotoxicity of ocular implants was developed using fluorescent viability dyes. To assess the assay’s sensitivity in detecting toxic substances on biomaterials, zinc diethydithiocarbamate (ZDEC) and benzalkonium chloride (BAK) were deposited on silicone surfaces at different concentrations. Human lens epithelial cells (HLEC) were added to the surface of these treated silicone surfaces and were assessed for viability. The viability of both the adherent and non-adherent cells was determined using confocal microscopy with, annexin V, ethidium homodimer, and calcein. Cell metabolism was also evaluated using resazurin and the release of inflammatory cytokines was quantified using a multiplex Mesoscale Discovery platform. Confocal microscopy was shown to be a sensitive assay for evaluating material toxicity, as significant toxicity (p < 0.05) from ZDEC and BAK-treated surfaces compared to the untreated silicone control was detected. Patterns of cytokine release from cells varied depending on the toxin evaluated and the toxin concentration and did not directly correlate with the reduction in cell metabolic activity measured by alamarBlue.

authors

  • Kao, Elizabeth CY
  • Seo, Junghee
  • McCanna, David J
  • Subbaraman, Lakshman N
  • Jones, Lyndon

publication date

  • March 17, 2022