The performance evaluation of direct detection electron energy-loss spectroscopy at 200 kV and 80 kV accelerating voltages
Abstract
Direct electron detectors (DeDs) have been widely used for imaging studies because of their higher beam sensitivity, lower noise, improved pixel resolution, etc. However, there have been limited studies related to the performance in spectroscopic applications for the direct electron detection. Hereby, taking the advantage of the DeD installed on a high-performance electron energy-loss spectrometer, we systematically studied the performance of a …