Home
Scholarly Works
High-Resolution MR Venography at 3.0 Tesla
Journal article

High-Resolution MR Venography at 3.0 Tesla

Abstract

PURPOSE: The aim of this study was to investigate the visualization of small venous vessels in the normal human brain at a field strength of 3 Tesla. METHODS: T2*-weighted, three-dimensional gradient-echo images were acquired by exploiting the magnetic susceptibility difference between oxygenated and deoxygenated hemoglobin in the vasculature and microvasculature. The spatial resolution was 0.5 x 0.5 x 1 mm3, and sequence parameters were varied to obtain good vessel delineation. Improved visibility of venous vessels was obtained by creating phase mask images from the magnetic resonance phase images and multiplying these by the magnitude images. Venograms were created by performing a minimum intensity projection over targeted volumes. RESULTS: Highly detailed visualization of venous structures deep in the brain and in the superficial cortical areas were obtained without administration of an exogenous contrast agent; compared with similar studies performed at 1.5 T, the echo time could be reduced from typically 40-50 ms to 17-28 ms. CONCLUSION: Imaging at high-field strength offers the possibility of improved resolution and the delineation of smaller vessels compared with lower field strengths.

Authors

Reichenbach JR; Barth M; Haacke EM; Klarhöfer M; Kaiser WA; Moser E

Journal

Journal of Computer Assisted Tomography, Vol. 24, No. 6, pp. 949–957

Publisher

Wolters Kluwer

Publication Date

November 1, 2000

DOI

10.1097/00004728-200011000-00023

ISSN

0363-8715

Contact the Experts team