Magnetic susceptibility measurements of ultrathin films using the surface magneto-optic Kerr effect: Optimization of the signal-to-noise ratio Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • The magnetic susceptibility, χ=∂M/∂H, can be determined for ultrathin films using the surface magneto-optic Kerr effect and an ac technique (ac-MOKE) where the response to a small, modulated H field is measured. Optimization of the signal-to-noise ratio is imperative if the measurements are to be used to extract critical scaling exponents from the small tail of χ in the paramagnetic region. These optimal conditions are different than those appropriate for more conventional static H measurements (dc-MOKE) used to generate magnetic hysteresis loops, because the relative noise contributions from the laser stability, sample stability, and detector are changed. A simple method for determining optimum conditions is demonstrated, and the technique is used to show that a 1.8 ML Fe/W(110) film has the critical exponent γ of the 2D Ising model.

publication date

  • November 1, 1997