Journal article
In-vivo and in-vitro measurements of lead and cadmium
Abstract
Tibia lead is measured in vivo using X-ray fluorescence. A109Cd source is used to excite Pb K X-rays, and this signals is normalized to that from Rayleigh scattering to remove geometrical variations. The lower limit of detection is 10 μg/g for a mean absorbed dose, to the exposed section of the leg, of 100 μGy. Tibia lead correlated positively with age in normal volunteers (r=0.615,p=0.004) and with duration of exposure in occupationally …
Authors
Chettle DR; Franklin DM; Guthrie CJG; Scott MC; Somervaille LJ
Journal
Biological Trace Element Research, Vol. 13, No. 1, pp. 191–208
Publisher
Springer Nature
Publication Date
August 1987
DOI
10.1007/bf02796632
ISSN
0163-4984