Home
Scholarly Works
Effects of incomplete decay in fluorescence...
Journal article

Effects of incomplete decay in fluorescence lifetime estimation

Abstract

Fluorescence lifetime imaging has emerged as an important microscopy technique, where high repetition rate lasers are the primary light sources. As fluorescence lifetime becomes comparable to intervals between consecutive excitation pulses, incomplete fluorescence decay from previous pulses can superimpose onto the subsequent decay measurements. Using a mathematical model, the incomplete decay effect has been shown to lead to overestimation of the amplitude average lifetime except in mono-exponential decays. An inverse model is then developed to correct the error from this effect and the theoretical simulations are tested by experimental results.

Authors

Leung RWK; Yeh S-CA; Fang Q

Journal

Biomedical Optics Express, Vol. 2, No. 9, pp. 2517–2531

Publisher

Optica Publishing Group

Publication Date

September 1, 2011

DOI

10.1364/boe.2.002517

ISSN

2156-7085

Contact the Experts team