The reproducibility of 109Cd-based X-ray fluorescence measurements of bone lead. Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • We assessed the reproducibility of X-ray fluorescence-based lead measurements from multiple measurements made on a low-concentration plaster of paris phantom and in five subjects measured five times on two occasions. Over a 6-month period, 220 measurements of the same phantom were obtained and showed a standard deviation of 1.29 micrograms Pb (g plaster of paris)-1. The two sets of in vivo measurements were made 10 months apart and revealed a mean standard deviation of 3.4 micrograms Pb (g bone mineral)-1 and 5.1 micrograms Pb (g bone mineral)-1 for males and females, respectively. Our measured standard deviation exceeded by 20-30% the calculated standard deviation associated with a single measurement both in the phantom and in subjects. This indicates that some variance is introduced during the measurement process. Operator learning and consistency significantly minimized this increased variability. Measured lead concentrations of the left and right tibia in 14 subjects showed no significant differences between legs. As a result, either tibia can be sampled and compared over time. The levels of reproducibility we report here mean that X-ray fluorescence-based determinations of bone lead concentrations are reliable both over the short and long term. Thus, reasonably sized confidence intervals can be placed on detected changes in concentration and should permit acquisition of longitudinal data within a reasonable length of time.

publication date

  • August 1994